Loading [MathJax]/extensions/MathMenu.js
Primary Realization of Inductance and Capacitance Scales With a Fully Digital Bridge | IEEE Journals & Magazine | IEEE Xplore

Primary Realization of Inductance and Capacitance Scales With a Fully Digital Bridge


Abstract:

This article describes an automated electronic fully digital bridge for the comparison of four-terminal-pair (4TP) impedance standards in the audio frequency range. The b...Show More

Abstract:

This article describes an automated electronic fully digital bridge for the comparison of four-terminal-pair (4TP) impedance standards in the audio frequency range. The bridge relative accuracy, which is on the order of 10-6, makes it suitable as a reference bridge for the realization of primary scales of inductance and capacitance in metrology institutes and calibration laboratories. The performances of this bridge were validated by comparing the results of the calibrations of inductance and capacitance standards with those obtained from an existing analog reference system based on the three-voltage method. The article also reports the results of this validation.
Article Sequence Number: 1503008
Date of Publication: 13 October 2022

ISSN Information:

Funding Agency:


References

References is not available for this document.