Abstract:
This paper presents a multimodal multichannel neural activity readout IC which can perform not only the electrical recording (ER) but also the fluorescence recording (FR)...Show MoreMetadata
Abstract:
This paper presents a multimodal multichannel neural activity readout IC which can perform not only the electrical recording (ER) but also the fluorescence recording (FR) of neural activity for the cell-type-specific study of heterogeneous neuronal cell populations. The time-based FR circuit senses Ca2+ concentration using Ca2+ probes while the ER circuit acquires action potentials (APs) and local field potentials (LFPs). The IC is fabricated in 0.18μm CMOS. The FR circuit achieves a recording range of 81dB (75pA to 860nA) and consumes the power of 0.7μmW/Ch. The ER circuit achieves the input-referred noise (IRN) of 2.7μm Vrms over the bandwidth (BW) of 10kHz, while consuming the power of 4.9μmW/Ch. The in-vitro measurement is performed for recording Ca2+ concentration and electrical neural signals.
Published in: 2019 Symposium on VLSI Circuits
Date of Conference: 09-14 June 2019
Date Added to IEEE Xplore: 29 July 2019
ISBN Information:
ISSN Information:
School of EE, KAIST, Daejeon, Korea
UNIST, Ulsan, Korea
UNIST, Ulsan, Korea
KIST, Seoul, Korea
School of EE, KAIST, Daejeon, Korea
Ewha Womans University, Seoul, Korea
KIST, Seoul, Korea
UNIST, Ulsan, Korea
School of EE, KAIST, Daejeon, Korea
School of EE, KAIST, Daejeon, Korea
UNIST, Ulsan, Korea
UNIST, Ulsan, Korea
KIST, Seoul, Korea
School of EE, KAIST, Daejeon, Korea
Ewha Womans University, Seoul, Korea
KIST, Seoul, Korea
UNIST, Ulsan, Korea
School of EE, KAIST, Daejeon, Korea