Calibration of a LCR-meter at arbitrary phase angles using a fully automated impedance simulator


Abstract:

A fully automated impedance simulator is used to calibrate a commercial LCR-meter over the whole complex plane. The amplitude of the synthesized impedance has been varied...Show More

Abstract:

A fully automated impedance simulator is used to calibrate a commercial LCR-meter over the whole complex plane. The amplitude of the synthesized impedance has been varied over 5 orders of magnitude i.e. from 100 Ω to more than 10 MΩ and the phase angle from -90 deg to +90 deg. The first measurements have been carried out at a frequency of 1 kHz, however, the bandwidth of the instrument is 20 kHz.
Date of Conference: 10-15 July 2016
Date Added to IEEE Xplore: 11 August 2016
ISBN Information:
Electronic ISSN: 2160-0171
Conference Location: Ottawa, ON, Canada