Abstract:
A simple calibration method, which determines inductance from the capacitance standard using a commercially available LCR meter, is presented. Two 10 mH standard inductor...Show MoreMetadata
Abstract:
A simple calibration method, which determines inductance from the capacitance standard using a commercially available LCR meter, is presented. Two 10 mH standard inductors were calibrated against a 1 muF reference capacitor at 1592 Hz. The relative combined standard uncertainty was estimated to be 7.4 muH/H
Date of Conference: 27 June 2004 - 02 July 2004
Date Added to IEEE Xplore: 12 February 2007
ISBN Information: