Abstract:
A universal calibration method for four-terminal-pair (4TP) admittance standards, developed for frequencies up to around 10 MHz, is presented. The method is applicable to...Show MoreMetadata
Abstract:
A universal calibration method for four-terminal-pair (4TP) admittance standards, developed for frequencies up to around 10 MHz, is presented. The method is applicable to calibration of conversion adapters that convert a system from a 4TP to either a three-terminal system or a one-port system. Implementation of the method requires only a set of driving-point-impedance measurements, using proven one-port technology, and calculations of 4TP transfer admittance and output impedances using an impedance matrix.<>
Published in: Conference on Precision Electromagnetic Measurements
Date of Conference: 11-14 June 1990
Date Added to IEEE Xplore: 06 August 2002