Abstract:
In this paper, a novel transmitting coil for magnetic resonance wireless power transfer is designed, which is called honeycomb-shape coil. It surrounds the human body and...Show MoreMetadata
Abstract:
In this paper, a novel transmitting coil for magnetic resonance wireless power transfer is designed, which is called honeycomb-shape coil. It surrounds the human body and can form a stronger magnetic field than the traditional coil. Take the leadless cardiac pacemaker as an example, suppose it is on the left side of the heart and design the LCC-C compensation topology circuit to build an energy transmission system working at 13.56MHz, the simulation results show that the transmission efficiency is approximately 0.9%. When the received power is 1mW, the maximum average specific absorption rate (SAR) of the system is 0.240W/kg, which accords with the EN62209-1 human electromagnetic radiation standard.
Published in: 2022 International Applied Computational Electromagnetics Society Symposium (ACES-China)
Date of Conference: 09-12 December 2022
Date Added to IEEE Xplore: 20 March 2023
ISBN Information:
Funding Agency:
References is not available for this document.