Abstract:
The authors have developed a ferrocene-mediated amperometric glucose oxidase enzyme electrode for continuous in vivo measurement of subcutaneous tissue glucose in man. Th...Show MoreMetadata
First Page of the Article

Abstract:
The authors have developed a ferrocene-mediated amperometric glucose oxidase enzyme electrode for continuous in vivo measurement of subcutaneous tissue glucose in man. This sensor can be used to detect hypoglycemia. In vitro studies have shown that the response time of the sensor is less than one minute and probably does not contribute significantly to the observed phase lag between the nadir for the blood and subcutaneous glucose. In the subjects studied to date, the mean delay was about 3 min, but the range was 0-15 minutes.<>
Published in: Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society
Date of Conference: 04-07 November 1988
Date Added to IEEE Xplore: 10 April 2006
Print ISBN:0-7803-0785-2