Abstract:
The relationship between surface diaphragm mechanomyography (sMMGdi), as a noninvasive measure of inspiratory muscle mechanical activation, and crural diaphragm electromy...Show MoreMetadata
Abstract:
The relationship between surface diaphragm mechanomyography (sMMGdi), as a noninvasive measure of inspiratory muscle mechanical activation, and crural diaphragm electromyography (oesEMGdi), as the invasive gold standard measure of diaphragm electrical activation, had not previously been examined. To investigate this relationship, oesEMGdi and sMMGdi were measured simultaneously in 6 healthy subjects during an incremental inspiratory threshold loading protocol, and analyzed using fixed sample entropy (fSampEn). A positive curvilinear relationship was observed between mean fSampEn sMMGdi and oesEMGdi (r = 0.67). Accordingly, an increasing electromechanical ratio was also observed with increasing inspiratory load. These findings suggest that sMMGdi could provide useful noninvasive measures of inspiratory muscle mechanical activation.
Published in: 2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
Date of Conference: 18-21 July 2018
Date Added to IEEE Xplore: 28 October 2018
ISBN Information:
ISSN Information:
PubMed ID: 30441104