Abstract:
The development of ambulatory arterial pulse oximetry is key to longer term monitoring and treatment of cardiovascular and respiratory conditions. The investigation prese...Show MoreMetadata
Abstract:
The development of ambulatory arterial pulse oximetry is key to longer term monitoring and treatment of cardiovascular and respiratory conditions. The investigation presented in this paper will assist the designer of an ambulatory pulse oximetry monitor in minimizing the overall LED power consumption (PLED,TOT ) levels by analyzing the lowest achievable limit as constrained by the optical components, circuitry implementation and final SpO2 reading accuracy required. LED duty cycle (DLED ) reduction and light power (PLED,TOT ) minimization are proposed as methods to reduce p . Bandwidth and signal quality calculations are carried out in order to determine the required PLED,TOT as a function of the different noise sources.
Published in: 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society
Date of Conference: 22-26 August 2007
Date Added to IEEE Xplore: 22 October 2007
ISBN Information:
ISSN Information:
PubMed ID: 18002450