Abstract:
We confirm possibility of noninvasive measurements of hemoglobin concentration by means of wrist wearable device. The dataset consists of 170 volunteers with the referenc...Show MoreMetadata
Abstract:
We confirm possibility of noninvasive measurements of hemoglobin concentration by means of wrist wearable device. The dataset consists of 170 volunteers with the reference hemoglobin concentrations in the range of 65–176 g/L. The developed algorithm predicts hemoglobin concentration with the mean absolute error as low as 6.8 g/L and correlation coefficient with the reference blood test equal 0.66.
Date of Conference: 04-08 December 2023
Date Added to IEEE Xplore: 26 February 2024
ISBN Information: