An ex-vivo rat sciatic nerve preparation was developed to quantify activation thresholds of fiber subtypes and electrode material charge injection limits for platinum iri...
Impact Statement:This study establishes a repeatable experimental paradigm in which to characterize electrochemical performance of novel electrode materials in the context of nerve activa...Show More

Abstract:

Emerging therapies in bioelectronic medicine highlight the need for deeper understanding of electrode material performance in the context of tissue stimulation. Electroch...Show More
Impact Statement:
This study establishes a repeatable experimental paradigm in which to characterize electrochemical performance of novel electrode materials in the context of nerve activation for bioelectronic medical applications.

Abstract:

Emerging therapies in bioelectronic medicine highlight the need for deeper understanding of electrode material performance in the context of tissue stimulation. Electrochemical properties are characterized on the benchtop, facilitating standardization across experiments. On-nerve electrochemistry differs from benchtop characterization and the relationship between electrochemical performance and nerve activation thresholds are not commonly established. This relationship is important in understanding differences between electrical stimulation requirements and electrode performance. We report functional electrochemistry as a follow-up to benchtop testing, describing a novel experimental approach for evaluating on-nerve electrochemical performance in the context of nerve activation. An ex-vivo rat sciatic nerve preparation was developed to quantify activation thresholds of fiber subtypes and electrode material charge injection limits for platinum iridium, iridium oxide, titanium nitride and PEDOT. Finally, we address experimental complexities arising in these studies, and demonstrate statistical solutions that support rigorous material performance comparisons for decision making in neural interface development.
An ex-vivo rat sciatic nerve preparation was developed to quantify activation thresholds of fiber subtypes and electrode material charge injection limits for platinum iri...
Page(s): 59 - 65
Date of Publication: 22 January 2024
Electronic ISSN: 2644-1276
PubMed ID: 38445242