Abstract:
Some issues of the direct Fourier method (DFM) implementation are discussed. A hybrid spline-linear interpolation for the DFM is proposed. The results of comprehensive si...Show MoreMetadata
First Page of the Article

Abstract:
Some issues of the direct Fourier method (DFM) implementation are discussed. A hybrid spline-linear interpolation for the DFM is proposed. The results of comprehensive simulation research are presented. The following reconstruction problems and parameters are emphasized: interpolation, increasing the radial density of the polar raster, filtering, the 2-D inverse Fourier transformation dimension, and considering the cases of noiseless and noisy input data. For the a priori prescribed resolution of the reconstructed image, values of reconstruction parameters have been determined which are optimal with regard to reconstruction quality and computation cost. The computational requirements of the DFM algorithm which correspond to distinct interpolation schemes are compared to one another for CT and MR tomography, respectively. The estimations obtained are compared to computational characteristics of the convolution backprojection method.<>
Published in: IEEE Transactions on Medical Imaging ( Volume: 9, Issue: 4, December 1990)
DOI: 10.1109/42.61757