Thermal metrology of an electrically active silicon heated atomic force microscope cantilever and doped polysilicon microbeams was performed using Raman spectroscopy. The temperature dependence of the Stokes Raman peak location and the Stokes to anti-Stokes intensity ratio calibrated the measurements, and it was possible to assess both temperature and thermal stress behavior with resolution near 1mum. The devices can exceed 400degC with the required power depending upon thermal boundary conditions. Comparing the Stokes shift method to the intensity ratio technique, non-negligible errors in devices with mechanically fixed boundary conditions compared to freely standing structures arise due to thermally induced stress. Experimental values were compared with a finite element model, and were within 9% of the thermal response and 5% of the electrical response across the entire range measured
Published in:
Components and Packaging Technologies, IEEE Transactions on
(Volume:30
,
Issue:
2
)
Date of Publication: June 2007