The effect of 1.75-D array element height on time-shift compensation is analyzed using a two-dimensional array system and a distributed aberration phantom. The analysis uses hydrophone measurements of transmit beams. The results indicate that, for imaging at 3 MHz through aberration representative of the abdominal wall, similar focus compensation performance can be expected from arrays with element pitches less than or equal to 3.0 mm.
Published in:
Ultrasonics, Ferroelectrics and Frequency Control, IEEE Transactions on
(Volume:49
,
Issue:
9
)
Date of Publication: Sept. 2002