International comparison for RF power in the frequency range up to 18 GHz

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

Formats Non-Member Member
$31 $13
Learn how you can qualify for the best price for this item!
Become an IEEE Member or Subscribe to
IEEE Xplore for exclusive pricing!
close button

puzzle piece

IEEE membership options for an individual and IEEE Xplore subscriptions for an organization offer the most affordable access to essential journal articles, conference papers, standards, eBooks, and eLearning courses.

Learn more about:

IEEE membership

IEEE Xplore subscriptions

15 Author(s)
de Vreede, J.P.M. ; NMi Van Swinden Lab., Delft, Netherlands ; Korfage, W. ; Persson, P. ; Brunetti, L.
more authors

A Euromet international laboratory comparison (Project 393) has been carried out between 14 national standards laboratories. Thermistor mounts were used, equipped with PC7 as well as Type N connectors. The comparison is carried out using the normal equipment of the laboratory for high-quality external calibration. The results show good agreement in measuring the calibration coefficient of the thermistor mounts within the claimed expanded uncertainty (typically between 1% and 2%). It confirms the equivalence of national standards for RF power up to 18 GHz. In one case, corrective action is proposed

Published in:

Instrumentation and Measurement, IEEE Transactions on  (Volume:50 ,  Issue: 2 )