Reduction of false blood pressure alarms by use of electrocardiogram blood pressure relationships

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

Formats Non-Member Member
$31 $13
Learn how you can qualify for the best price for this item!
Become an IEEE Member or Subscribe to
IEEE Xplore for exclusive pricing!
close button

puzzle piece

IEEE membership options for an individual and IEEE Xplore subscriptions for an organization offer the most affordable access to essential journal articles, conference papers, standards, eBooks, and eLearning courses.

Learn more about:

IEEE membership

IEEE Xplore subscriptions

3 Author(s)
Zong, W. ; Div. of Health Sci. & Technol., MIT, Cambridge, MA, USA ; Moody, G.B. ; Mark, R.G.

This paper presents an algorithm for reducing false alarms related to changes in arterial blood pressure (ABP) in intensive care unit (ICU) monitoring. The algorithm assesses the ABP signal quality, analyzes the ECG-ABP relationships using a fuzzy logic approach, and post-processes (accepts or rejects) ABP alarms produced by a commercial monitor. The algorithm was developed and evaluated using disjoint sets of data from the MIMIC database. By rejecting 96.8% of the false ABP alarms (182 of 188) produced by the monitor using the test set of data, the algorithm was able to reduce the false alarm rate from 34.9% to 1.7% of ABP alarms while accepting 99.4% of true ABP alarms (346 of 348). The results show that the algorithm is both effective and practical, and feasible for use in future monitoring systems

Published in:

Computers in Cardiology, 1999

Date of Conference:

1999