Calibration of a 3 MV LI measuring system by comparison method and by numerical convolution using its measured step response

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

Formats Non-Member Member
$31 $31
Learn how you can qualify for the best price for this item!
Become an IEEE Member or Subscribe to
IEEE Xplore for exclusive pricing!
close button

puzzle piece

IEEE membership options for an individual and IEEE Xplore subscriptions for an organization offer the most affordable access to essential journal articles, conference papers, standards, eBooks, and eLearning courses.

Learn more about:

IEEE membership

IEEE Xplore subscriptions

2 Author(s)
Breilmann, W. ; Darmstadt Univ. of Technol., Germany ; Haupt, H.

The dynamic errors of peak value uˆ, front time T1 and time to half value T2 of an approved 3 MV LI (lightning impulse) measuring system were experimentally determined for two standard LIs with uˆ=500 kV, T1=1.2 μs±20% and T2=60 μs by two different methods. The first method is according to the comparison of an Approved Measuring System with a 500 kV Reference Measuring System. The second method is the convolution of the normalized step response g(t) of the 3 MV system with the standard LI input voltage measured with the Reference Measuring System. Agreement within 0.5% for uˆ and within 1.2% for T1 and T2 was achieved