Evaluation of low energy threshold settings for PVI PET systems

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

Formats Non-Member Member
$31 $13
Learn how you can qualify for the best price for this item!
Become an IEEE Member or Subscribe to
IEEE Xplore for exclusive pricing!
close button

puzzle piece

IEEE membership options for an individual and IEEE Xplore subscriptions for an organization offer the most affordable access to essential journal articles, conference papers, standards, eBooks, and eLearning courses.

Learn more about:

IEEE membership

IEEE Xplore subscriptions

3 Author(s)
Kohlmyer, S.G. ; Div. of Nucl. Med., Washington Univ., Seattle, WA, USA ; Miyaoka, R.M. ; Lewellen, T.K.

The authors studied the effects of adjusting the low level energy discriminator (LLD) for positron volume imaging (PVI) mode acquisitions using a GE Advance PET system. NEMA scatter fraction and count loss measurements were performed for a 20 cm right circular cylinder at LLD between 300 keV and 450 keV. From these data, noise equivalent count rate (NECR) curves were calculated to estimate effects of LLD on scans of head sized objects. To evaluate the effects of LLD on whole-body image quality, contrast-to-noise ratio (CNR) values were obtained from images processed without attenuation or scatter correction. CNR data were acquired for 1.6 cm and 1.0 cm spherical lesions within the liver region of a Data Spectrum torso phantom scanned with arms, and activity outside the field of view. The results indicate that the NECR for a head-sized object is fairly insensitive to LLD differences between 300 keV and 425 keV for activity concentrations up to 0.4 /spl mu/Ci/cc. The NECR improved slightly for higher concentrations with 375 keV

Published in:

Nuclear Science, IEEE Transactions on  (Volume:46 ,  Issue: 6 )