Highnam, R. ; Dept. of Eng. Sci., Oxford Univ., UK ; Brady, M. ; English, R.

Microcalcifications can be one of the earliest signs of breast cancer. Unfortunately, their appearance in mammograms can be mimicked by dust and dirt entering the imaging process and this has been shown previously to lead to false positives. The authors use a model of the imaging process and, in particular, the blurring functions inherent within it to detect the film-screen artifacts caused by dust and dirt and, thus, reduce false-positives. A crucial facet of the work is the choice of the correct image representation upon which to perform the image processing. After extensive testing, the authors' algorithm has identified no microcalcifications as being artifacts and has an artifact detection rate of approaching 96%.

Published in: