The thermoelectric transfer difference of Thermal Converters (TCs) can be evaluated in both current mode and voltage mode, using a fast-reversed dc (FRDC) source. Some types of TCs show differences in the thermoelectric transfer difference between voltage and current mode as large as some parts in 106. In order to investigate the origin of the discrepancy, the change of the thermoelectric transfer difference with the value of a resistor in series with the thermal converter has been measured. The result was in good agreement with an assumption that the Seebeck effect in the heater/heater support junction is the main source of the voltage transfer difference
Published in:
Instrumentation and Measurement, IEEE Transactions on
(Volume:48
,
Issue:
2
)
Date of Publication: Apr 1999