Carrillo, A. ; Dept. of Biomed. Eng., Case Western Reserve Univ., Cleveland, OH, USA ; Wilson, D.L.

We are developing technology for thermal ablation of tumors under interventional MRI guidance. Image registration allows us to quantitatively and visually compare images acquired before, during and after treatment. One manual and two automatic registration methods were evaluated using displacement of anatomical landmarks on clinical and volunteer studies. Manual registration gave an RMS distance error of around 4 mm on points in the middle of the liver. This is sufficiently small to allow the use of image registration in the evaluation of thermal tumor ablation in the liver

Published in:

Engineering in Medicine and Biology Society, 1997. Proceedings of the 19th Annual International Conference of the IEEE  (Volume:2 )

Date of Conference:

30 Oct-2 Nov 1997