Ultrasound doppler measurements of low velocity blood flow: limitations due to clutter signals from vibrating muscles

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

Formats Non-Member Member
$31 $13
Learn how you can qualify for the best price for this item!
Become an IEEE Member or Subscribe to
IEEE Xplore for exclusive pricing!
close button

puzzle piece

IEEE membership options for an individual and IEEE Xplore subscriptions for an organization offer the most affordable access to essential journal articles, conference papers, standards, eBooks, and eLearning courses.

Learn more about:

IEEE membership

IEEE Xplore subscriptions

2 Author(s)
Heimdal, A. ; Dept. of Physiol. & Biomed. Eng., Norwegian Univ. of Sci. & Technol., Trondheim, Norway ; Torp, H.

Skeletal muscles vibrate under sustained contraction, and generate low frequency side band clutter in the doppler signal. Both shivering in the hand of the operator and muscle vibrations in the patient itself give rise to the clutter. Clutter rejection filters are commonly used to remove the low frequency components, but the doppler signal from low velocity blood flow is then also lost. This paper describes a model for the pulsed wave (PW) doppler signal from vibrating muscles, reviews a model for the PW doppler signal from moving blood, and by comparing these models presents a theoretical minimum for detectable blood velocity in small vessels, being typically 6.4 mm/s for 6 MHz doppler. The limit has a nonlinear relation to the ultrasound frequency. The model also shows that the radial component of the muscle vibrations can be estimated from the phase of the doppler signal.

Published in:

Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on  (Volume:44 ,  Issue: 4 )