A dynamic calibration technique for temperature programmed desorption spectroscopy

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

The purchase and pricing options are temporarily unavailable. Please try again later.
6 Author(s)
Hurst, K. E. ; National Renewable Energy Laboratory, 15013 Denver West Parkway, Golden, Colorado 80403, USA ; Heben, M. J. ; Blackburn, J. L. ; Gennett, T.
more authors

Your organization might have access to this article on the publisher's site. To check, click on this link:http://dx.doi.org/+10.1063/1.4770115 

A novel, rapid and accurate calibration procedure as a means for quantitative gas desorption measurement by temperature programmed desorption (TPD) spectroscopy is presented. Quantitative measurement beyond the linear regime of the instrument is achieved by associating an instantaneous calibrated molar flow rate of gas to the detector response. This technique is based on fundamental methods, and is independently verified by comparison to the hydrogen desorption capacity of a known standard metal hydride with known stoichiometry. The TPD calibration procedure described here may be used for any pure gas, and the accuracy is demonstrated for the specific case of hydrogen.

Published in:

Review of Scientific Instruments  (Volume:84 ,  Issue: 2 )