In developing world health clinics, incidence of surgical site infection is 2 to 10 times higher than in developed world hospitals. This paper identifies lack of availability of appropriately designed, low-cost autoclaves in developing world health clinics as a major contributing factor to the dramatic gap in surgical site infection rates. The paper describes the process of developing a low-cost autoclave that addresses the unique challenges faced by developing world primary health clinics and discusses how appropriateness of design was determined. The resulting pressure cooker-based autoclave design was fabricated and tested against the CDC specifications. Twelve partnering clinics in Nepal trialed these autoclaves from July until December 2012.
Published in:
Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE
Date of Conference: Aug. 28 2012-Sept. 1 2012