Induced voltage to an implantable cardiac pacemaker in a voxcel human model by a near-field intra-body communication device

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

Formats Non-Member Member
$31 $13
Learn how you can qualify for the best price for this item!
Become an IEEE Member or Subscribe to
IEEE Xplore for exclusive pricing!
close button

puzzle piece

IEEE membership options for an individual and IEEE Xplore subscriptions for an organization offer the most affordable access to essential journal articles, conference papers, standards, eBooks, and eLearning courses.

Learn more about:

IEEE membership

IEEE Xplore subscriptions

6 Author(s)
Yoshino, Y. ; Dept. of Electr. & Electron. Eng., Tokyo Metropolitan Univ., Tokyo, Japan ; Taki, M. ; Furuya, A. ; Shibata, N.
more authors

The interference voltage at the terminals of a unipolar-type implantable cardiac pacemaker induced by a near-field intra-body communication device was investigated by numerical calculations. Operating frequency was assumed 5 MHz. The dependency of induced voltage on the electrical constants of a realistic human model was examined. The results showed that the induced voltage was inversely proportional to the absolute value of the complex relative permittivity. The induced voltage was much smaller than the immunity requirement on interference voltage for active medical implants when the applied input voltage to the communication device was within the practical use condition.

Published in:

General Assembly and Scientific Symposium, 2011 XXXth URSI

Date of Conference:

13-20 Aug. 2011