Correction of spatial distortion, gain nonuniformity and efficiency variation in detectors for positron emission mammography

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

Formats Non-Member Member
$31 $13
Learn how you can qualify for the best price for this item!
Become an IEEE Member or Subscribe to
IEEE Xplore for exclusive pricing!
close button

puzzle piece

IEEE membership options for an individual and IEEE Xplore subscriptions for an organization offer the most affordable access to essential journal articles, conference papers, standards, eBooks, and eLearning courses.

Learn more about:

IEEE membership

IEEE Xplore subscriptions

5 Author(s)
Robar, J.L. ; Med. Phys. Unit, McGill Univ., Montreal, Que., Canada ; Thompson, C.J. ; Murthy, K. ; Clancy, R.L.
more authors

Each of two detectors used in our positron emission mammography (PEM) system consists of four 36 mm×36 mm×20 mm bismuth germanate (BGO) crystal detector blocks coupled to a crossed-wire anode position-sensitive photomultiplier tube (PS-PMT). To facilitate high spatial-resolution imaging, the crystal blocks have been pixellated into small elements on two opposite faces. In total, there are 71×71 elements on each detector. In order to improve the quality of images obtained using this system, we have identified three required aspects of detector calibration. First, in order to correct for spatial distortion inherent in the detectors, an accurate method for crystal element identification has been developed. This algorithm successfully identifies 59×49 crystal elements, and these results are used to generate a distortion Look-Up-Table (LUT). Second, corrections for a marked spatial variation of detector gain are required in order to facilitate accurate energy discrimination. This variation is quantitated by generating energy spectra specific to each of the identified crystal elements, and recording the relative gain of each element in an energy LUT. Finally, by using a third LUT containing the relative efficiencies of crystal elements, the flood-image nonuniformity has been reduced from 50% to 13%

Published in:
Nuclear Science Symposium, 1996. Conference Record., 1996 IEEE  (Volume:2 )

Date of Conference: 2-9 Nov 1996

Need Help?


IEEE Advancing Technology for Humanity About IEEE Xplore | Contact | Help | Terms of Use | Nondiscrimination Policy | Site Map | Privacy & Opting Out of Cookies

A not-for-profit organization, IEEE is the world's largest professional association for the advancement of technology.
© Copyright 2013 IEEE - All rights reserved. Use of this web site signifies your agreement to the terms and conditions.