Phase displacement correction of reference channel in digital bridge used for calibation of instrument transformers with digital output

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

Formats Non-Member Member
$33 $13
Learn how you can qualify for the best price for this item!
Become an IEEE Member or Subscribe to
IEEE Xplore for exclusive pricing!
close button

puzzle piece

IEEE membership options for an individual and IEEE Xplore subscriptions for an organization offer the most affordable access to essential journal articles, conference papers, standards, eBooks, and eLearning courses.

Learn more about:

IEEE membership

IEEE Xplore subscriptions

1 Author(s)
J. I. Juvik ; SP National Testing and Research Institute of Sweden, P.O Box 857, SE-501 15 Borås, Sweden

Phase displacements in the reference channel of a digital bridge used for calibrating instrument transformers with digital output have been investigated. Since the sampling in this set-up is controlled from the unit under test the largest correction stems from the virtual time shift caused by the aperture time integration when measuring a sinusoidal voltage with the sampling multimeter in DCV mode. This time shift is found to be equal to half the aperture time. In addition is the correction for the reaction time of the multimeter in this mode found to be 140 ns.