Toward Simultaneous 2D ACAR and 2D DBAR: Sub-Pixel Spatial Characterization of a Segmented HPGe Detector Using Transient Charges

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

The purchase and pricing options are temporarily unavailable. Please try again later.
5 Author(s)
Williams, C.S. ; Air Force Inst. of Technol., Wright-Patterson AFB, OH, USA ; Baker, W.P. ; Burggraf, Larry W. ; Adamson, P.E.
more authors

Sub-pixel spatial characterization of an Ortec HPGe double-sided strip detector was measured in preparation for its planned use in a positron annihilation spectrometer (PAS) to simultaneously measure the two-dimensional Angular Correlation (2D ACAR) and Doppler Broadening (2D DBAR) of Annihilation Radiation. Sr-85 514-keV photons were finely collimated onto the center of 25, equally spaced sub-pixels within a single 5-mm by 5-mm intrinsic pixel, using a collimator with an aperture diameter of 0.15 ?? 0.05 mm. The location of each full-energy recorded event was determined by analyzing the relative sizes of coincident transient charges on charge collection strips adjacent to the strip recording full-energy events. Interpolation, using ratios of the coincident transient charge sizes, was used to determine two-dimensional locations of full-energy events. Radial location data was fit to a function which describes a Gaussian point spread function uniformly distributed over a circular aperture. The standard deviation of the point spread function, 0.22 mm, is a measure of the spatial resolution of this detector system. The relative efficiency for detection of 514-keV photons across the intrinsic pixel was also measured.

Published in:

Nuclear Science, IEEE Transactions on  (Volume:57 ,  Issue: 2 )