Your organization might have access to this article on the publisher's site. To check, click on this link:http://dx.doi.org/+10.1063/1.3293459
A simple method to prepare gold tips for scanning tunneling microscopy has been introduced. In this method, electrochemical etching without a hazardous electrolyte was employed. The setup uses basic laboratory instrumentation to control the etching process. This avoids purchasing complicated, expensive, and dedicated equipment for tip preparation. A procedure to optimize the etching parameters by setting the current limit is described. Etched tips were checked with both optical and scanning electron microscopy.
Published in:
Review of Scientific Instruments
(Volume:81
,
Issue:
1
)
Date of Publication: Jan 2010