Influence of temperature on the response of diamond radiation detectors

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

Formats Non-Member Member
$31 $31
Learn how you can qualify for the best price for this item!
Become an IEEE Member or Subscribe to
IEEE Xplore for exclusive pricing!
close button

puzzle piece

IEEE membership options for an individual and IEEE Xplore subscriptions for an organization offer the most affordable access to essential journal articles, conference papers, standards, eBooks, and eLearning courses.

Learn more about:

IEEE membership

IEEE Xplore subscriptions

7 Author(s)
Tromson, D. ; LETI (CEA–Technologies Avancées)/DEIN/SPE, CEA/Saclay, 91191 Gif-sur-Yvette, France ; Brambilla, A. ; Bergonzo, P. ; Mas, A.
more authors

Your organization might have access to this article on the publisher's site. To check, click on this link:http://dx.doi.org/+10.1063/1.1347411 

The response of chemically vapor deposited diamond detectors under alpha particle irradiation was studied as a function of operating temperature in the 20–280 °C range. The detector characteristics were studied for two initial states: (i) the detector being either preirradiated under x-rays at a dose of 8 Gy (pumped state) or (ii) preannealed at 400 °C for 15 min (depumped state). Pumped detectors were found to exhibit a stable response (evaluated by measurement of the charge collection efficiency η) with temperature up to 120 °C, above which a decrease of η was observed. Preannealing at 400 °C was shown to induce, at room temperature, a drastic deterioration of the detector response, when compared to that of pumped devices, whereas above 150 °C a reversed trend was observed. Correlations between the response of the diamond detectors and the presence of traps were determined by comparing the detector response for α particles with the thermally stimulated current (TSC) spectrum. The temperature dependence of the detector response was shown to be related to the presence of a trapping level measured around 260 °C by TSC. © 2001 American Institute of Physics.

Published in:

Journal of Applied Physics  (Volume:90 ,  Issue: 3 )

Date of Publication:

Aug 2001

Need Help?


IEEE Advancing Technology for Humanity About IEEE Xplore | Contact | Help | Terms of Use | Nondiscrimination Policy | Site Map | Privacy & Opting Out of Cookies

A not-for-profit organization, IEEE is the world's largest professional association for the advancement of technology.
© Copyright 2014 IEEE - All rights reserved. Use of this web site signifies your agreement to the terms and conditions.