Spectral resolution measurement of an x-ray imaging crystal spectrometer for KSTAR

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

The purchase and pricing options are temporarily unavailable. Please try again later.
7 Author(s)
Lee, S.G. ; National Fusion Research Institute, Daejeon 305-333, Republic of Korea ; Bak, J.G. ; Nam, U.W. ; Moon, M.K.
more authors

Your organization might have access to this article on the publisher's site. To check, click on this link:http://dx.doi.org/+10.1063/1.2982416 

A spectral resolution measurement of the x-ray imaging crystal spectrometer for the Korea Superconducting Tokamak Advanced Research machine utilizing a segmented position-sensitive detector and time-to-digital converter based delay-line readout electronics was performed by using an x-ray tube in a laboratory. The measured spectral resolution is about 12 600, which means the actual energy resolution is 0.32 eV for the x-ray tube’s bremsstrahlung peak energy of 4 keV. The results from the spectral resolution measurement are described.

Published in:

Review of Scientific Instruments  (Volume:79 ,  Issue: 10 )