Spring constant calibration of atomic force microscopy cantilevers with a piezosensor transfer standard

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

Formats Non-Member Member
$31 $31
Learn how you can qualify for the best price for this item!
Become an IEEE Member or Subscribe to
IEEE Xplore for exclusive pricing!
close button

puzzle piece

IEEE membership options for an individual and IEEE Xplore subscriptions for an organization offer the most affordable access to essential journal articles, conference papers, standards, eBooks, and eLearning courses.

Learn more about:

IEEE membership

IEEE Xplore subscriptions

5 Author(s)
Langlois, E.D. ; Materials Science & Engineering Laboratory, National Institute of Standards & Technology, Boulder, Colorado 80305, USA and Department of Mechanical Engineering, University of Colorado, Boulder, Colorado 80309, USA ; Shaw, G.A. ; Kramar, J.A. ; Pratt, J.R.
more authors

Your organization might have access to this article on the publisher's site. To check, click on this link:http://dx.doi.org/+10.1063/1.2785413 

We describe a method to calibrate the spring constants of cantilevers for atomic force microscopy (AFM). The method makes use of a “piezosensor” composed of a piezoresistive cantilever and accompanying electronics. The piezosensor was calibrated before use with an absolute force standard, the NIST electrostatic force balance (EFB). In this way, the piezosensor acts as a force transfer standard traceable to the International System of Units. Seven single-crystal silicon cantilevers with rectangular geometries and nominal spring constants from 0.2 to 40 N/m were measured with the piezosensor method. The values obtained for the spring constant were compared to measurements by four other techniques: the thermal noise method, the Sader method, force loading by a calibrated nanoindentation load cell, and direct calibration by force loading with the EFB. Results from different methods for the same cantilever were generally in agreement, but differed by up to 300% from nominal values. When used properly, the piezosensor approach provides spring-constant values that are accurate to ±10% or better. Methods such as this will improve the ability to extract quantitative information from AFM methods.

Published in:

Review of Scientific Instruments  (Volume:78 ,  Issue: 9 )