High speed tapping mode atomic force microscopy in liquid using an insulated piezoelectric cantilever

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

The purchase and pricing options are temporarily unavailable. Please try again later.
11 Author(s)
Rogers, B. ; Department of Mechanical Engineering and the Nevada Ventures Nanoscience Program, University of Nevada, Reno, Reno, Nevada 89557 ; Sulchek, T. ; Murray, K. ; York, D.
more authors

Your organization might have access to this article on the publisher's site. To check, click on this link:http://dx.doi.org/+10.1063/1.1619548 

Quicker imaging times for tapping mode atomic force microscopy in liquid could provide a real-time imaging tool for studying dynamic phenomena in physiological conditions. We demonstrate faster imaging speed using microcantilevers with integrated piezoelectric actuators. The exposed electric components of the cantilever necessitate an insulation scheme for use in liquid; three coating schemes have been tested. Preliminary tapping mode images have been taken using the insulated microactuator to simultaneously vibrate and actuate the cantilever over topographical features in liquid, including a high speed image of steps on a mica surface in water and an image of two e coli bacteria taken in saline solution at 75.5 μm/s, a threefold improvement in bandwidth versus conventional piezotube actuators. © 2003 American Institute of Physics.

Published in:

Review of Scientific Instruments  (Volume:74 ,  Issue: 11 )