Continuum x-ray source as a calibration system for charge coupled devices

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

The purchase and pricing options are temporarily unavailable. Please try again later.
13 Author(s)
Hashimotodani, K. ; Department of Earth and Space Science, Graduate School of Science, Osaka University, 1-1, Machikaneyama-cho, Toyonaka, Osaka 560, Japan ; Toneri, T. ; Kitamoto, S. ; Tsunemi, H.
more authors

Your organization might have access to this article on the publisher's site. To check, click on this link:http://dx.doi.org/+10.1063/1.1148671 

We present x-ray spectra obtained by a charge coupled device (CCD) from an electron impacting x-ray source using several kinds of electron targets: Au, Al–bronze, Al, and pure Al. The x rays from the source are dispersed by a grating and the dispersed x rays are focused on the CCD. Owing to the fine spatial and moderate energy resolutions of the CCD, fine spectra with 2 eV energy resolution up to 2.2 keV are obtained. X rays from the pure Al target (99.999% purity) provide good continuum x rays except for Al K x rays. This continuum source is useful as a calibration source for low energy x-ray detectors. © 1998 American Institute of Physics.

Published in:

Review of Scientific Instruments  (Volume:69 ,  Issue: 2 )