In situ force calibration of high force constant atomic force microscope cantilevers

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

The purchase and pricing options are temporarily unavailable. Please try again later.
3 Author(s)
Scholl, D. ; Research Staff, Ford Motor Company, MD‐3028, Dearborn, Michigan 48121‐2053 ; Everson, M.P. ; Jaklevic, R.C.

Your organization might have access to this article on the publisher's site. To check, click on this link:http://dx.doi.org/+10.1063/1.1144737 

A miniature capacitive force sensor which fits in the sample position of an atomic force microscope (AFM) has been used to calibrate the force applied by the scanning tip during nanohardness measurements. The sensor is simple, physically robust, and easily fabricated from readily available materials. It can be operated with commonly available electronic instrumentation. The device described here is optimized for tip‐sample forces between 10-4 and 10-2 N, a range useful for micrometer‐scale mechanical modification of surfaces. With minor design changes, sensors of this general type should be capable of much greater sensitivity. The sensor is calibrated outside the AFM by the application of free weights. Calculated estimates of AFM cantilever force constants are difficult for all but the simplest geometries and may rely on uncertain values for certain critical dimensions and material properties. These estimates can thereby be replaced by a simple and direct measurement.

Published in:

Review of Scientific Instruments  (Volume:65 ,  Issue: 7 )