Your organization might have access to this article on the publisher's site. To check, click on this link:http://dx.doi.org/+10.1063/1.1139570
An apparatus for the rotation of sample and blank in diffuse reflectance Fourier transform infrared spectroscopy has been designed and fabricated. Spectra taken with sample rotation give data with better precision by a factor of 4 than without rotation. Additionally, sample rotation saves time by eliminating the need for taking three to five spectra followed by spectrum averaging as has been done in the past to obtain a representative spectrum of solids.
Published in:
Review of Scientific Instruments
(Volume:58
,
Issue:
6
)
Date of Publication: Jun 1987