By Topic

Optimized linear variable differential transformer displacement transducer for cryogenic use

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

The purchase and pricing options are temporarily unavailable. Please try again later.
2 Author(s)
Sponagle, N.C. ; Department of Physics, Queen''s University, Kingston, Ontario, Canada K7L 3N6 ; Atherton, D.L.

Your organization might have access to this article on the publisher's site. To check, click on this link:http://dx.doi.org/+10.1063/1.1136307 

A simple procedure is given for optimizing the performance and coil dimensions of an Ellis and Walstrom type moving coil, linear variable differential transformer displacement transducer as well as a method of compensating for changes in calibration at different operating temperatures. The resulting transducer has been used successfully in liquid He and ambient magnetic fields up to 3 T but may also be used at any temperature up to 300 K without recalibration.

Published in: