Tip motion in amplitude modulation (tapping-mode) atomic-force microscopy: Comparison between continuous and point-mass models

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

The purchase and pricing options are temporarily unavailable. Please try again later.
2 Author(s)
Rodrı guez, Tomas R. ; Instituto de Microelectrónica de Madrid, CSIC, Isaac Newton 8, 28760 Tres Cantos, Madrid, Spain ; Garcı a, Ricardo

Your organization might have access to this article on the publisher's site. To check, click on this link:http://dx.doi.org/+10.1063/1.1456543 

We discuss the influence of high-order frequency components in the operation of an amplitude modulation atomic-force microscope (AFM). A comparative study of point-mass and continuous models is performed to describe the tip motion. The tip–surface interaction force excites high-order frequency components whenever a higher harmonic of the excitation force is close to an eigenmode of the cantilever beam. The strength of those components depends on the set point amplitude and the fundamental resonance frequency of the cantilever. However, for standard operating conditions with quality factors in the 102–103 range, higher-order components are about three orders of magnitude smaller than the component at the excitation frequency. We conclude that point-mass models are suitable to describe the operation of a tapping-mode AFM in air environments. © 2002 American Institute of Physics.

Published in:

Applied Physics Letters  (Volume:80 ,  Issue: 9 )