The scatter background in high-resolution pinhole SPECT

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

Formats Non-Member Member
$31 $13
Learn how you can qualify for the best price for this item!
Become an IEEE Member or Subscribe to
IEEE Xplore for exclusive pricing!
close button

puzzle piece

IEEE membership options for an individual and IEEE Xplore subscriptions for an organization offer the most affordable access to essential journal articles, conference papers, standards, eBooks, and eLearning courses.

Learn more about:

IEEE membership

IEEE Xplore subscriptions

3 Author(s)
McKee, B.T.A. ; Dept. of Radiol., Ottawa Univ., Ont., Canada ; Jammal, R.B. ; Chamberlain, M.J.

The authors have evaluated the scattered background in high-resolution pinhole SPECT by Monte Carlo modeling of a point source of 140 keV photons with an 8 cm diameter water-equivalent sphere as a scattering object. The detector geometry was 5 cm between center and pinhole, and 15 cm from pinhole to detector. The pinhole diameter ranged from 2 to 4 mm. For comparison, the authors have also modeled the same source and scatter geometry for a high-resolution parallel-hole SPECT detector. The detected scattered/unscattered ratio is about 0.2. This ratio and the shape of the scattered background are almost constant for both pinhole and parallel-hole geometry and over a range of pinhole diameters. The shape, though not the amplitude, of the scattered background is independent of axial motion of the source. The pinhole geometry imposes some distortion from scattering at a distance from the central plane. Some implications of these modelling results for scatter correction in pinhole SPECT are discussed

Published in:

Nuclear Science Symposium and Medical Imaging Conference, 1994., 1994 IEEE Conference Record  (Volume:4 )

Date of Conference:

30 Oct-5 Nov 1994