A comparison of 3D scatter correction methods in a neuroPET tomograph

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

Formats Non-Member Member
$31 $13
Learn how you can qualify for the best price for this item!
Become an IEEE Member or Subscribe to
IEEE Xplore for exclusive pricing!
close button

puzzle piece

IEEE membership options for an individual and IEEE Xplore subscriptions for an organization offer the most affordable access to essential journal articles, conference papers, standards, eBooks, and eLearning courses.

Learn more about:

IEEE membership

IEEE Xplore subscriptions

4 Author(s)
Spinks, T.J. ; Clinical Sci. Centre, Hammersmith Hospital, London, UK ; Grootoonk, S. ; Bailey, D.L. ; Schnorr, L.

A comparison has been made between two procedures for scatter correction of 3D PET data acquired on a brain scanner (CTI/Siemens 953B). The methods are (i) dual energy window (DW), in which a weighted subtraction of data in a scatter window (200-380 keV) is made from the photopeak (380-850 keV) and (ii) convolution-subtraction (CS), in which a space-invariant kernel is derived from line source data. In multi-compartment phantoms, contrast ratios were restored to within about 6% of their true values with similar results for DW and CS. For both methods, good correction was observed in inactive water-filled compartments but residual counts were observed in air-filled cavities. Data from a patient injected with FDG was obtained by acquiring a 3D frame immediately after a standard 2D dynamic acquisition. Comparison of 2D and 3D images (scaled in kBq/ml using cross-calibration factors derived from phantoms) was made by plotting corresponding 8 mm square ROIs over the whole brain. The slopes/intercepts of these plots (ideally 1.0 and 0) were 0.97/-0.008 (DW), 0.95/-0.006 (CS) and 1.18/0.028 (uncorrected). Thus, overall similar results were obtained for the two methods although the noise in the corrected data was slightly higher for DW than for CS

Published in:

Nuclear Science Symposium and Medical Imaging Conference, 1994., 1994 IEEE Conference Record  (Volume:4 )

Date of Conference:

30 Oct-5 Nov 1994