Signal Versus Noise Characteristics of Filtered EMG Used as a Control Source

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

Formats Non-Member Member
$31 $13
Learn how you can qualify for the best price for this item!
Become an IEEE Member or Subscribe to
IEEE Xplore for exclusive pricing!
close button

puzzle piece

IEEE membership options for an individual and IEEE Xplore subscriptions for an organization offer the most affordable access to essential journal articles, conference papers, standards, eBooks, and eLearning courses.

Learn more about:

IEEE membership

IEEE Xplore subscriptions

1 Author(s)
Kreifeldt, J.G. ; Department of Engineering Graphics and Design, Tufts University, Medford, Mass.

A common method of initially processing surface-detected electromographic (EMG) activity is to differentially amplify, rectify, and then smooth (using a low-pass filter) the rectified activity. The output of the filter exhibits a mean value slightly obscured by ripple. Both the mean value and ripple increase with increasing muscle contraction. The mean value can be called the signal while the ripple, undesirable when only the signal is desired, can be called noise. The SNR depends, at least, upon the contraction level, type of smoothing filter, and the amount of smoothing for the particular filter. This defined SNR is important in signal communication problems of both a design and a theoretical nature.

Published in:

Biomedical Engineering, IEEE Transactions on  (Volume:BME-18 ,  Issue: 1 )