Electrolysis-a factor in cardiac pacemaker electrode failure

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

The purchase and pricing options are temporarily unavailable. Please try again later.
1 Author(s)

During continuous pacing experiments at the University of Missouri Medical Center, Columbia, with the micromodule pacemaker receiver' implanted on the left ventricle of thirty-six mongrel dogs in complete heart block, it was observed that after 24 days in one case, 52 days in another, and 35 days in a third, micromodules with stainless steel electrodes failed to pace. Upon reoperation it was discovered that the positive electrode of one module had been reduced to a short stub. In another module the positive electrode was loose, disrupting the circuit. In the third module the positive electrode was completely missing.

Published in:

Bio-medical Electronics, IEEE Transactions on  (Volume:10 ,  Issue: 4 )

Date of Publication:

Oct. 1963