The influence of different electrode geometries on the current density distribution during radio-frequency ablation. A three-dimensional finite element study

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

The purchase and pricing options are temporarily unavailable. Please try again later.
5 Author(s)
Panescu, D. ; EP Technol. Inc., Sunnyvale, CA, USA ; Fleishman, S.D. ; Whayne, J.G. ; Swanson, D.K.
more authors

The locations of current density maxima can provide information about the locations of temperature maxima. Our goals were: (1) to find the optimal place of the temperature sensor used for temperature-controlled ablation; (2) to assess at which locations char is more likely to occur. Using finite element modeling, we analyzed the current density distribution for a 8Fr/5 mm straight and a 8Fr/8 mm valve ablation electrodes. The maximal current densities occurred at the tip of the electrodes. Hence, it is best to place the temperature sensor, which provides input for the temperature control, at the electrode tip. Also, the analysis showed that using an electrically insulating coating layer at the junction between the electrode and the catheter body significantly reduced the local current density, therefore the local temperature. This avoids the formation of charring

Published in:

Engineering in Medicine and Biology Society, 1994. Engineering Advances: New Opportunities for Biomedical Engineers. Proceedings of the 16th Annual International Conference of the IEEE

Date of Conference:

1994