Design and performance evaluation of a new high resolution detector array module for PET

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

Formats Non-Member Member
$31 $13
Learn how you can qualify for the best price for this item!
Become an IEEE Member or Subscribe to
IEEE Xplore for exclusive pricing!
close button

puzzle piece

IEEE membership options for an individual and IEEE Xplore subscriptions for an organization offer the most affordable access to essential journal articles, conference papers, standards, eBooks, and eLearning courses.

Learn more about:

IEEE membership

IEEE Xplore subscriptions

2 Author(s)
Tzanakos, G. ; Dept. of Phys., Athens Univ., Greece ; Pavlopoulos, S.

The authors have designed a 2-D detector-array module for high resolution PET. The module consists of a 16×18 matrix of crystals cut from a single BGO block coupled to a 3"×3" square envelope position-sensitive PMT. The performance of the detector module was studied using simulation techniques and the design was optimized to provide high resolution, good energy and timing characteristics, and ease in the construction. The energy resolution of the module at the photopeak was estimated to 27%, whereas the timing resolution of a coincident pair of detector modules was found to be 6.5 ns FWHM

Published in:

Nuclear Science Symposium and Medical Imaging Conference, 1993., 1993 IEEE Conference Record.

Date of Conference:

31 Oct-6 Nov 1993