Signal-to-noise ratios for attenuation correction in PET imaging

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

Formats Non-Member Member
$31 $13
Learn how you can qualify for the best price for this item!
Become an IEEE Member or Subscribe to
IEEE Xplore for exclusive pricing!
close button

puzzle piece

IEEE membership options for an individual and IEEE Xplore subscriptions for an organization offer the most affordable access to essential journal articles, conference papers, standards, eBooks, and eLearning courses.

Learn more about:

IEEE membership

IEEE Xplore subscriptions

3 Author(s)
McKee, B.T.A. ; Dept. of Phys., Queen''s Univ., Kingston, Ont., Canada ; Hiltz, L.G. ; Harvey, P.J.

Attenuation correction is an important part of accurate image reconstruction in positron tomography. The usual correction method involves direct measurement of attenuation correction factors (ACFs). A reconstruct-reproject method, which has been suggested as providing superior noise properties, is sometimes employed; an attenuation image is first reconstructed from the measurement and then ACFs are obtained by reprojection through this image. Here the authors present a model which follows the signal-to-noise ratio (SNR) through the attenuation correction by both the direct and reconstruct-reproject methods. This model is applicable to both 2D and 3D imaging geometry, but applies to the central elements of emission and transmission objects with circular symmetry and constant amplitude. For this simplified geometry, the model predicts that the SNR of the emission image following attenuation correction is the same for both direct and reconstruct-reproject methods, although the SNRs of the ACFs are themselves substantially different. The authors also present the measured SNR at the various steps of attenuation correction for both the direct and reconstruct-reproject methods using simulated transmission and emission data. The measured SNRs agree with the model; no significant difference between the direct and reconstruct-reproject SNRs was observed

Published in:

Medical Imaging, IEEE Transactions on  (Volume:13 ,  Issue: 4 )