5 Author(s)
Yemelyanov, S.Y. ; Inst. of Math. Problems of Biol., Acad. of Sci., Pushchino, Russia ; Skovoroda, A.R. ; Lubinski, M.A. ; Shapo, B.M.
more authors

Quantitative strain images of tissue equivalent phantoms have been obtained for a number of conditions using Fourier-based speckle tracking methods. Experimental results demonstrate that a non-palpable inclusion with Young's modulus only 3 times greater than surrounding material can be easily detected. Comparison of experimental results with theoretical simulations quantifies the overall accuracy of the method

Published in:

Ultrasonics Symposium, 1992. Proceedings., IEEE 1992

Date of Conference:

20-23 Oct 1992