Two new strategies to increase the signal to noise ratio in positron volume imaging

Sign In

Cookies must be enabled to login.After enabling cookies , please use refresh or reload or ctrl+f5 on the browser for the login options.

Formats Non-Member Member
$33 $13
Learn how you can qualify for the best price for this item!
Become an IEEE Member or Subscribe to
IEEE Xplore for exclusive pricing!
close button

puzzle piece

IEEE membership options for an individual and IEEE Xplore subscriptions for an organization offer the most affordable access to essential journal articles, conference papers, standards, eBooks, and eLearning courses.

Learn more about:

IEEE membership

IEEE Xplore subscriptions

2 Author(s)
C. J. Thompson ; Montreal Neurol. Inst., McGill Univ., Montreal, Que., Canada ; Y. Picard

Two new hardware techniques which improve the scatter fraction of positron volume imaging (PVI) systems are presented. The first uses the fact that Compton scattered rays scattered through a large angle have a lower energy. The range of angles through which rays can be scattered and still be in the scanner field of view (FOV) depends on the radius of the line of response. By using a low-energy threshold which is proportional to the line of response's (LOR's) radius, the scatter fraction at the edges of the image is substantially reduced. The second technique is based on the observation that in NaI crystals many of the 511-keV rays are backscattered, and the scattered photon then escapes undetected. The use of a notched energy acceptance window, which includes the photo-peak and the backscatter peak, but discards the scattered events in between, is proposed

Published in:

IEEE Transactions on Nuclear Science  (Volume:40 ,  Issue: 4 )