This paper examines a novel approach for temporal calibration of a three-dimensional (3-D) freehand ultrasound system. A localization system fixed on the probe gives the position and orientation of the probe. For quantitative use, calibration is needed to correctly localize a B-scan in four-dimensional (4-D) (3-D+t) space. Temporal latency estimation is defined in a general robust formulation using no specific probe motion constraints. Experiments were performed on synthetic and real data using a 3-D freehand ultrasound system. The achieved precision is lower than the image acquisition rate (40 ms). A validation study using a calibration phantom has been performed to evaluate the influence of incorrect latency estimation on the 3-D reconstruction procedure. We showed that for latency estimation errors less than 40 ms, the 3-D reconstruction errors are negligible for volume estimation
Published in:
Medical Imaging, IEEE Transactions on
(Volume:25
,
Issue:
8
)
Date of Publication: Aug. 2006