Bai Baodong ; Sch. of Electr. Eng., Shenyang Univ. of Tech. ; Zhang Yanli ; Xie Dexin ; Xia Pingchou

A new unilateral permanent magnet configuration for magnetic resonance imaging (MRI) system producing a slice imaging region with field homogeneity is presented. It is a totally new problem about magnet design to build a slice imaging region with homogeneous magnetic field outside the magnet source. A novel design method with high precision is proposed here to probe the existence possibility of such a remote slice region and then is used to analyze the location and size of the slice region. The success of such a unilateral permanent magnet with a slice region for MRI builds a strong basis for some scientific research such as the construction of the unilateral MRI system with homogeneous magnetic field

Published in:

Engineering in Medicine and Biology Society, 2005. IEEE-EMBS 2005. 27th Annual International Conference of the

Date of Conference:

17-18 Jan. 2006