P. R. B. Marinho ; Lab. de Sistemas de Deteccao, Centro Brasileiro de Pesquisas Fisicas, Rio de Janeiro, Brazil ; A. F. Barbosa ; G. P. Guedes

The development of a triple-GEM imaging detector is reported. The triple-GEM detector provides two-dimensional (2-D) localization of ionizing events making use of a multilayer printed circuit board (X&Y anode) for position readout. Two-dimensional localization is achieved by associating one discrete delay line to each spatial coordinate. Results referring to differential nonlinearity (DNL), integral nonlinearity (INL) and spatial resolution are presented. The detector has been operated with Ar/Xe/CO2 (64/16/20) gas mixture at atmospheric pressure. 5.9 keV (55Fe source) photons were used to characterize the detector. The feasibility of the imaging technique is demonstrated and the major issues are addressed

Published in:

IEEE Transactions on Nuclear Science  (Volume:52 ,  Issue: 6 )